C9404 – (bizhub 364e/284e/224e)

You are here:

Contents

Trouble typeCIS LED lighting abnormally
Trouble codeC9404
RankA
Detection timing<When DF-701 is installed>
After a scan job is completed, the machine detected that the average output of the CIS LED is greater than the specified value.
Trouble isolationScanner
Relevant electrical parts• CIS module (CIS)
• CIS power supply (CISPU)
• Dual scan image processing board (DSIPB)
• MFP board (MFPB)

 (2)Procedure

StepActionControl signalLocation of electrical components
1Check the connection between SCDB-LU201, and correct if necessary.
2Check the connector between SCDB PJ2-PRCB CN43 for proper connection and correct as necessary.
3Check the connector between CCDB PJ3-MFPB PJ01 for proper connection and correct as necessary.
4Replace the LED exposure unit (Part# A61FM31100)
5Replace CCDB (Part# A61FR72100)
6SCDB F1 conduction check
7Replace SCDB (Part# A5C1H01A11)
8Replace PRCB (Part# A61FH00101)
9Replace MFPB (Part# A61FH02005)

Please search Knowledge Base articles and Bulletins as they may contain relevant information and updated Part information such as Part numbers.

K.3.13.1 (Link to the Field Service Manual)