C0216 (bizhub 368/308)
Contents
Trouble type | C0216: External LCT up/down abnormality (When LU-302 is installed) |
Rank | B |
Trouble detection condition | The LU upper limit sensor (PS2) is not turned ON (blocked) even after the set period of time has elapsed after the paper lift-up operation for the drawer began. |
Trouble isolation | LCT |
Relevant electrical parts | • LU upper limit sensor (PS2) • LU lift-up motor (M1) • LU drive board (LUDB) |
Procedure
- Check the connector between M1-LUDB CN3 for proper connection and correct as necessary.
- Check the connector of M1 for proper drive coupling and correct as necessary.
- Check the connector between PS2-relay CN3-LUDB CN5 for proper connection and correct as necessary.
- PS2 I/O check, sensor check
- Control signal: LUDB CN5-3 (ON)
- Location of electrical component: LU-302 4-G
- M1 operation check
- Control signal: LUDB CN3-4 (ON)
- Location of electrical component: LU-302 3-G
- Replace M1. (Part# A00JR74900)
- Replace LUDB. (Part# A87VH01000)
NOTICE
- Link to the wiring diagram (LU-302)
Please search Knowledge Base articles and Bulletins as they may contain relevant information and updated Part information such as Part numbers.
K.3.7.13 (Link to the Field Service Manual)